How do I perform an DQ IQ OQ PQ RQ using EasyMatch QC-ER?

Posted on Dec 03, 2025 by HunterLab

HunterLabのEasyMatch QC-ER(電子記録)ソフトウェアは、通常、製薬会社が透明および半透明の液体、半透明のクリーム、粉末、圧搾プラーク、丸い錠剤、楕円形のカプレットおよびカプセル、ならびに包装材料の色を測定するためのFDA 21 CFR Part 11要件を満たすために購入されます。

HunterLabは、このソフトウェアとともに、以下のIQOQPQプロトコルを実装したバリデーションおよびコンプライアンスノートブックを提供します:

IQ - 設置資格

IQ - Installation Qualification は、HunterLab装置がセットアップされ、電源が入れられ、必要に応じてソフトウェアがPCにインストールされ、装置の標準化が完了することでハードウェアとソフトウェアの両方が正しく動作することが確認されるプロトコルを定義しています。

バリデーション&コンプライアンス・ノートブックに続き、IQは以下の3つの要素で構成される:

  1. Procedure 1 - Installation of Hardware すべてのセンサーについて、センサーの電源が入り、準備が整ったところまでセンサーを設置します。センサーの通信ケーブルをPCに接続します。
  2. 手順2 - EZMQC-ERソフトウェアの全システムへのインストール.
  3. EasyMatch QC-ER が正常に開かれ、センサーがロードされます。センサーがビープ音を鳴らすか反応したら、ソフトウェアとセンサー間の電源と通信が確立されたことになります。装置の標準化に進みます。成功すると、IQ設置認定が達成されます。

OQ - オペレーション・クオリフィケーション

OQ - Operation Qualification は、センサー診断テストを実行し、機器が製造元の許容範囲内で正しく動作していることを示すPASSを検証するプロトコルを定義しています。

  • Procedure 3 - Checking Wavelength Accuracy ディディミウム・フィルターの検証(球面測定器のみ)。
  • 手順4 - すべてのセンサーの白タイルで短期反復性をチェックする。
  • 手順5 - すべてのセンサーの緑色のタイルで中間域の性能または長期繰返し精度をチェックする。

Posted in EasyMatch QC-ER

How do I perform an DQ IQ OQ PQ RQ using EasyMatch QC-ER?

Posted on Feb 09, 2024 by HunterLab

HunterLab’s EasyMatch QC-ER (Electronic Records) software is typically purchased by pharmaceutical companies to meet FDA 21 CFR Part 11 requirements for measuring the color of transparent and translucent liquids, translucent creams, powders, pressed plaques, round tablets, oval caplets and capsules as well as packaging materials.

With this software HunterLab supplies a Validation and Compliance Notebook that implements IQOQPQ protocols as follows:

IQ – Installation Qualification

IQ – Installation Qualification defines a protocol by which a HunterLab instrument is set up and turned on; software is installed on a PC as needed and both hardware and software elements verified to be working correctly together by completing a successful instrument standardization.

Following the Validation and Compliance Notebook, IQ consists of the following three elements:

  1. Procedure 1 – Installation of Hardware for all sensors sensor to the point such that the instrument is powered up and ready. Connect the sensor communication cable to the PC.
  2. Procedure 2 – Installation of EZMQC-ER Software for All Systems.
  3. EasyMatch QC-ER is successfully opened and the sensor loaded. When the sensor beeps or responds, power and communications between software and sensor have been established. Proceed to standardize the instrument. When successful, IQ Installation Qualification has been achieved.

OQ – Operation Qualification

OQ – Operation Qualification defines a protocol in which the sensor diagnostics tests are run and verified to PASS indicating that the instrument is operating correctly to manufacturer’s tolerances.

  • Procedure 3 – Checking Wavelength Accuracy verification on the Didymium Filter for sphere instruments only.
  • Procedure 4 – Checking Short Term Repeatability on a White Tile for all sensors.
  • Procedure 5 – Checking Mid-Range Performance or Long Term Repeatability on the Green Tile for all sensors.